Subject(s) - medicine , forehead , core (optical fiber) , core temperature , liquid crystal , anesthesia , degree (music) , nuclear medicine , surgery , radiology , optics , physics , acoustics
Summary The use of cutaneous liquid crystal thermometry (EZ Temp) as an estimate of core temperature during routine surgery was investigated in 20 patients. Seventeen per cent of the recordings made with the EZ Temp were more than 1°C different from oesophageal temperature. There was a poor correlation between EZ Temp values and both oesophageal and aural temperatures (r = 0.54 for both sites). We conclude that liquid crystal thermometry of the forehead is not sufficiently accurate to be used as an indicator of core temperature during routine surgery.