An assessment of the Dinamap 845
Author(s) -
HUTTON P.,
DYE J.,
PRYSROBERTS C.
Publication year - 1984
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1984.tb07240.x
Subject(s) - medicine , blood pressure , sodium nitroprusside , anesthesia , nitric oxide
Summary The accuracy of the Dinamap 845 automatic blood pressure recorder was assessed by comparing its own indirect determinations of blood pressure with direct intra‐arterial recordings. It was found that in the majority of cases it was capable of producing reliable trend information during anaesthesia. The instrument may not be able to interpret pressure signals from a patient with a severe dysrhythmia. It is probably an unsuitable monitor for use with very rapidly acting drugs such as sodium nitroprusside.