z-logo
Premium
Characterization of a pulsatile rotary total artificial heart
Author(s) -
Jurney Patrick L.,
Glynn Jeremy J.,
Dykan Igor V.,
Hagen Matthew W.,
Kaul Sanjiv,
Wampler Richard K.,
Hinds Monica T.,
Giraud George D.
Publication year - 2021
Publication title -
artificial organs
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.684
H-Index - 76
eISSN - 1525-1594
pISSN - 0160-564X
DOI - 10.1111/aor.13810
Subject(s) - pulsatile flow , artificial heart , characterization (materials science) , cardiology , biomedical engineering , medicine , engineering , computer science , materials science , nanotechnology
This article describes the properties and performance of a rotary total artificial heart (TAH) that produces inherently pulsatile flow. The hydraulic performance of the TAH was characterized using a mock circulatory loop to simulate four physiologically relevant conditions: baseline flow, increased flow, systemic hypertension, and pulmonary hypertension. The pump has a variable shuttle rate (beats per minute), percentage dwell time, and angular velocity on either side (revolutions per minute), which allows for full control of the flow rate and pulsatility over a range of healthy and pathologic pressures and flow rates. The end‐to‐end length and displacement volume of the TAH are 9.8 cm and 130 mL, respectively, allowing it to fit in smaller chest cavities including those of smaller adults and juvenile humans.