Switch between AAI and DDD mode pacing—What is the mechanism?
Author(s) -
Ząbek Andrzej,
Małecka Barbara,
NowosielskaZąbek Ewa,
Dębski Maciej,
Boczar Krzysztof,
Lelakowski Jacek
Publication year - 2019
Publication title -
annals of noninvasive electrocardiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.494
H-Index - 48
eISSN - 1542-474X
pISSN - 1082-720X
DOI - 10.1111/anec.12648
Subject(s) - medicine , cardiology , implantable cardioverter defibrillator , mechanism (biology) , epistemology , philosophy
The electrocardiogram (ECG) interpretation in patients with implantable pacemaker is often a perplexing problem. The difficulty in the device evaluation increases in the presence of novel timing cycles and additional functions. Authors describe a special function frequently encountered in Medtronic dual‐chamber pacemakers and implantable cardioverter‐defibrillator devices called managed ventricular pacing (MVP) and demonstrate its performance in the patient with undersensing episodes in ventricular channel. Intermittent ventricular undersensing in the device with MVP feature turned on caused repetitive mode switches between AAI and DDD mode. This report shows unexceptional occurrence of tricky ECG findings in patient with Medtronic dual‐chamber device.