Comparison of the semen analysis results obtained from two branded computer‐aided sperm analysis systems
Author(s) -
Hu Y. A.,
Lu J. C.,
Shao Y.,
Huang Y. F.,
Lü N. Q.
Publication year - 2013
Publication title -
andrologia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.633
H-Index - 59
eISSN - 1439-0272
pISSN - 0303-4569
DOI - 10.1111/and.12010
Subject(s) - semen analysis , andrology , sperm , semen , biology , medicine , infertility , genetics , pregnancy
Summary The study evaluated the comparability of two branded computer‐aided sperm analysis ( CASA ) systems commonly used in andrology laboratories in C hina. The same semen sample was analysed using two branded CASA systems ( WLJY ‐9000 and CFT ‐9200) by one well‐trained technician. Results of semen analysis obtained from two branded CASA systems were then compared. The accuracy of counting results of CASA systems was evaluated using latex bead solutions with known concentrations of (35 ± 5) × 10 6 ml −1 and (18 ± 2.5) × 10 6 ml −1 . There were significant differences in all parameters ( P < 0.01) except for LIN and WOB . The counting results of CFT ‐9200 were close to the standard solutions [(38.86 ± 3.79) × 10 6 ml −1 and (19.03 ± 1.99) × 10 6 ml −1 ], while those of WLJY ‐9000 were underestimated [(28.53 ± 2.06) × 10 6 ml −1 and (14.62 ± 0.95) × 10 6 ml −1 ]. But the coefficient of variation of WLJY ‐9000 was lower than that of CFT ‐9200 (7.22%, 6.50% vs. 9.82%, 10.46%). It is concluded that factors such as parameter settings and evaluation algorithms could significantly affect the results obtained from these two branded CASA systems. Great attention should also be paid to the quality control in semen analysis with CASA .