The laboratory performance of the enFLOW ® , buddy lite ™ and ThermoSens ® fluid warmers
Author(s) -
Kim H. J.,
Yoo S. M.,
Jung J. S.,
Lee S. H.,
Sun K.,
Son H. S.
Publication year - 2015
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/anae.12906
Subject(s) - saline , isotonic saline , medicine , zoology , nuclear medicine , anesthesia , biology
Summary We measured heating of isotonic saline by three fluid warmers in six experiments: saline at 5 °C or 20 °C delivered at 30, 50 or 100 ml.min −1 . At the three flow rates, the enFLOW ® , buddy lite ™ and ThermoSens ® systems heated 5 °C saline to mean ( SD ) temperatures of: 41.1 (0.5) °C, 37.7 (0.6) °C and 39.1 (0.6) °C; to 40.3 (0.8) °C, 33.9 (1.6) °C and 39.3 (0.7) °C; and to 37.1 (0.8) °C, 24.0 (1.3) °C and 37.6 (1.0) °C, respectively, p < 0.0001 for each experiment. The mean ( SD ) times taken to heat 5 °C saline were: 16.6 (1.7) s, 258.4 (58.9) s and 134.2 (79.6) s; 16.9 (1.8) s, 256.2 (62.2) s and 182.5 (74.5) s; and 21.5 (1.5) s, 275.9 (49.3) s and 313.5 (18.0) s, respectively, p < 0.0003 for each experiment. The results for saline at 20 °C were similar. The enFLOW system heated saline above 36 °C faster than the ThermoSens system, whereas the buddy lite often failed to achieve 36 °C.