z-logo
Premium
A laboratory comparison of the performance of the buddy lite™ and enFlow™ fluid warmers
Author(s) -
Bruells C. S.,
Bruells A. C.,
Rossaint R.,
Stoppe C.,
Schaelte G.,
Zoremba N.
Publication year - 2013
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/anae.12415
Subject(s) - medicine , saline , outflow , volumetric flow rate , inflow , flow (mathematics) , anesthesia , mechanics , meteorology , physics
Summary Maintenance of normothermia is crucial to avoid patient morbidity. Newly released fluid warming devices have become smaller in size, but this change might impair efficacy. We performed an evaluation of the buddy lite™ and enFlow™ fluid warmers. We measured inflow and outlet temperatures of the devices at flow rates between 25 and 100 ml.min −1 using saline at room temperature or cooled to 10 °C. At a flow rate of 25 ml.min −1 , the outlet temperature of the buddy lite was significantly higher than that of the enFlow (p < 0.0001), but at flow rates of 75 and 100 ml.min −1 , it was significantly lower (p < 0.0001). This pattern was the same for both room temperature and cooled saline. There was a significant drop in the temperature of saline along the length of a 1‐m outflow tube. We conclude that both devices provide effective fluid warming at a low flow rate, although the heating capability of the buddy lite is limited at high flow rates.