z-logo
Premium
Measurement of heat and moisture exchanger efficiency
Author(s) -
Chandler M.
Publication year - 2013
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/anae.12368
Subject(s) - humidifiers , moisture , medicine , heat exchanger , humidity , relative humidity , process engineering , environmental science , thermodynamics , environmental engineering , composite material , materials science , physics , engineering
Summary Deciding between a passive heat and moisture exchanger or active humidification depends upon the level of humidification that either will deliver. Published international standards dictate that active humidifiers should deliver a minimum humidity of 33 mg.l −1 ; however, no such requirement exists, for heat and moisture exchangers. Anaesthetists instead have to rely on information provided by manufacturers, which may not allow comparison of different devices and their clinical effectiveness. I suggest that measurement of humidification efficiency, being the percentage moisture returned and determined by measuring the temperature of the respired gases, should be mandated, and report a modification of the standard method that will allow this to be easily measured. In this study, different types of heat and moisture exchangers for adults, children and patients with a tracheostomy were tested. Adult and paediatric models lost between 6.5 mg.l −1 and 8.5 mg.l −1 moisture (corresponding to an efficiency of around 80%); however, the models designed for patients with a tracheostomy lost between 16 mg.l −1 and 18 mg.l −1 (60% efficiency). I propose that all heat and moisture exchangers should be tested in this manner and percentage efficiency reported to allow an informed choice between different types and models.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here