In Vitro Evaluation of Resistance and Warming Performance of a Small Blood Warmer on a Continuous Renal Replacement Therapy Circuit
Author(s) -
Roy JeanPhilippe,
Reagor James,
Klein Tom,
Morgan Jolyn,
Palazzolo Alisa,
Ratcliff Joseph,
Goldstein Stuart L
Publication year - 2020
Publication title -
therapeutic apheresis and dialysis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.415
H-Index - 53
eISSN - 1744-9987
pISSN - 1744-9979
DOI - 10.1111/1744-9987.13349
Subject(s) - medicine , extracorporeal , hypothermia , venous return curve , anesthesia , blood flow , extracorporeal circulation , hemodynamics , surgery , cardiology
Children on extracorporeal devices are at increased risk of hypothermia and require circuits with lower extracorporeal volume to avoid blood priming. We performed an in vitro study of the enFlow blood warmer to assess its warming performance and impact on circuit resistance at a high blood flow rate. The enFlow was added on the return line of a continuous renal replacement therapy circuit in a closed circuit primed with expired packed red blood cells (40% Hct). Return venous pressure and temperature pre‐ and post‐ both enFlow and Prismaflo II were measured at varying blood flow with different combination of the two blood warmers. Each variable was recorded eight times over 2 min. Return venous pressure was higher with the enFlow but was still within clinically usable range (127.5 ± 4.6 vs. 45.3 ± 1.4 mm Hg at 200 mL/min, P  < 0.05). The enFlow was able to achieve post‐warmer temperature between 34.1 and 35.2°C even at 200 mL/min with blood, lower than what was observed with crystalloid, but still more effective than the Prismaflo II ( P  < 0.05 vs. enFlow post‐warmer temperature) achieving between 32.8 and 34.0°C. Combining both warmers achieved a higher post‐warmer temperature between 35.1 and 36.5°C. The enFlow can warm blood at high flow rates with minimal extracorporeal volume increase (~5 mL) and improve hypothermia prevention but may have resistance issues at the highest flows. The use of this device could benefit pediatric practice where smaller extracorporeal volumes are needed and patients are at increased risk of hypothermia.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here