Propagation‐based phase‐contrast X‐ray microtomography of a cerebral protection device retrieved after carotid artery stenting
Author(s) -
Youn Sung Won,
Kim Ho Kyun,
Kim Hong Tae,
Han Sung Mi,
Do Jin Kuk,
Do Young Rok,
Lee Hui Joong,
Lee Jongmin,
Lim Jae Hong
Publication year - 2014
Publication title -
journal of synchrotron radiation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.172
H-Index - 99
ISSN - 1600-5775
DOI - 10.1107/s1600577513023862
Subject(s) - materials science , carotid arteries , debris , medicine , stenosis , x ray microtomography , synchrotron , volume (thermodynamics) , radiology , nuclear medicine , biomedical engineering , optics , geology , physics , surgery , oceanography , quantum mechanics
Phase‐contrast synchrotron X‐ray microtomography (pcSyncX) based on the highly coherent X‐ray beam has previously been used to visualize the microstructures of biologic specimens, but it has never been used to evaluate embolic debris adherent on a cerebral protection device (CPD). The purpose of this study was to demonstrate the feasibility of pcSyncX for evaluating embolic debris during carotid artery stenting (CAS). Five patients (four males, age range 67–77 years) with severe carotid artery stenosis underwent CAS. The retrieved CPD was exposed to synchrotron radiation and 1000 pcSyncX projection images were obtained by rotating the CPD through 180°. An X‐ray shadow of a CPD was converted into a visual image by the scintillator. After microtomographic reconstruction, the three‐dimensionally reconstructed images were further segmented into the embolic debris and CPD. The total volume of emboli was calculated by summing the volume at each scanning level. The number of membrane pores covered by emboli as seen from the outer surface was counted and the percentage of covered area was calculated. Embolic debris was clearly demonstrated not only on the inner surface and within pores but also on the outer surface of the CPD. The mean total volume of embolic debris was 0.538 × 10 –6  mm 3 (range 0.225–0.965 × 10 –6  mm 3 ). Most (61.5%) of the debris was located at the apical one‐third of the CPD and 20.8% of the pore area was covered by debris.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here