A 2 m inelastic X‐ray scattering spectrometer at CMC‐XOR, Advanced Photon Source
Author(s) -
Hill J. P.,
Coburn D. S.,
Kim Y.J.,
Gog T.,
Casa D. M.,
Kodituwakku C. N.,
Sinn H.
Publication year - 2007
Publication title -
journal of synchrotron radiation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.172
H-Index - 99
ISSN - 1600-5775
DOI - 10.1107/s0909049507018006
Subject(s) - monochromator , scattering , optics , spectrometer , physics , photon , inelastic scattering , resolution (logic) , x ray , atomic physics , computer science , wavelength , artificial intelligence
The design and commissioning of an inelastic X‐ray scattering instrument at CMC‐XOR at the Advanced Photon Source is reported. The instrument features a 2 m vertical‐scattering arm with a novel counterweight design to reduce the twisting moment as the arm is moved in the scattering plane. A Ge(733) spherical analyzer was fabricated and an overall resolution of 118 meV (FWHM) was obtained with a Si(444) monochromator and a Si(111) pre‐monochromator. Early results from a representative cuprate, La 2 CuO 4 , are reported.