Development of a real‐time timing‐shutter performance monitor for protein crystallography
Author(s) -
Lee John,
Alkire R. W.,
Molitsky Michael,
Rotella F. J.,
Duke N. E. C.,
De Lurgio Patrick M.,
Madden Tim
Publication year - 2006
Publication title -
journal of synchrotron radiation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.172
H-Index - 99
ISSN - 1600-5775
DOI - 10.1107/s0909049506027610
Subject(s) - shutter , goniometer , rolling shutter , optics , photodiode , materials science , computer science , optoelectronics , physics
In order to accurately monitor shutter timing events and long‐term shutter performance, a timing‐shutter monitor has been developed. This monitor uses a photodiode to capture X‐ray‐induced fluorescence from the shutter blade in synchrony with goniometer rotation to measure shutter opening and closing delay times, as well as the total time that X‐rays are exposed to the sample during crystallographic data frames.