z-logo
open-access-imgOpen Access
A multiwire proportional counter for XAS fluorescence experiments
Author(s) -
Barrea Raúl A.,
Tamura E.,
Tolentino H. C. N.
Publication year - 2001
Publication title -
journal of synchrotron radiation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.172
H-Index - 99
ISSN - 1600-5775
DOI - 10.1107/s0909049500014746
Subject(s) - proportional counter , scintillator , physics , absorption (acoustics) , fluorescence , scintillation counter , detector , optics , geiger counter , absorption rate , analytical chemistry (journal) , materials science , chemistry , chromatography
A multiwire proportional counter was used in fluorescence X‐ray absorption measurements and a comparison to a Si(Li) and NaI(Tl) detectors was done. The main features of the mutiwire proportional counter are its high counting rate capability (10 7 counts s −1 ) and large active area (6 × 6 cm 2 ). It was shown that the MWPC is suitable for fluorescence absorption. Although the maximum capability was not reached in the present experiments, it was found that as the counting rate increase the MWPC performance became better than Si(Li) detectors and shows a similar response to the scintillator counter at medium counting rates (up to 10 5 counts s −1 ).

The content you want is available to Zendy users.