Premium
Ionization chambers for monitoring the position and tilt of X‐ray beams
Author(s) -
Arndt U. W.,
Kyte M. P.
Publication year - 2000
Publication title -
journal of applied crystallography
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.429
H-Index - 162
ISSN - 1600-5767
DOI - 10.1107/s0021889800000844
Subject(s) - collimated light , ionization , ionization chamber , tilt (camera) , beam (structure) , optics , anode , x ray , physics , position (finance) , materials science , atomic physics , ion , electrode , laser , mechanical engineering , finance , quantum mechanics , economics , engineering
An air‐filled ionization chamber with a three‐way split anode can be used to measure the tilt and the translation of a collimated X‐ray beam with respect to the chamber axis.