Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems
Author(s) -
Sofie L Hansen,
Natalie A. Bebbington
Publication year - 2020
Publication title -
nuclear medicine communications
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.489
H-Index - 73
eISSN - 1473-5628
pISSN - 0143-3636
DOI - 10.1097/mnm.0000000000001312
Subject(s) - pelvis , abdomen , thorax (insect anatomy) , siemens , nuclear medicine , medicine , reference range , image quality , radiology , computer science , physics , anatomy , artificial intelligence , quantum mechanics , image (mathematics) , endocrinology
CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocols prior to software version VB10 and a 75 kg adult thereafter. Quality reference mAs conversion factors are estimated for delivering equivalent mAs to children between two comparable SPECT-CT systems using adult and child references for topogram-based patient-size-related dose level adaptations.