Early unexpected failure of a vitamin E-infused highly cross-linked polyethylene liner
Author(s) -
Kyeong Baek Kim,
Sangmin Lee,
Nam Hoon Moon,
Min Uk,
Won Chul Shin
Publication year - 2021
Publication title -
medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.59
H-Index - 148
eISSN - 1536-5964
pISSN - 0025-7974
DOI - 10.1097/md.0000000000027454
Subject(s) - medicine , radiography , surgery , emergency department , psychiatry
Rationale: Total hip arthroplasty (THA) with a polyethylene (PE) liner is 1 of the most effective and successful treatment strategies for end-stage hip disorders. Vitamin E-infused highly cross-linked polyethylene is theoretically known to prevent failure due to oxidative degradation in the body, and is resistant to wear; therefore, successful long-term survival of THA is expected. Patient concerns: In June 2019, approximately 1.5 years after THA, the patient sat down and stood up without any specific issue; however, an unusual bullet sound occurred around the left hip joint. Since then, his discomfort persisted, and he was admitted to the emergency department. Diagnosis: Plain radiography and metal artifact reduction computed tomography performed in the emergency department revealed eccentric elevation of the prosthetic femoral head and suspected PE liner failure. Intervention: Revision surgery was performed for modular component exchange. To reduce the dislocation risk without performing cup exchange, conversion to dual-mobility articulation was performed. Outcomes: During the regular follow-up post-surgery, the patient could perform daily life activities without any discomfort, and dislocation was not observed. At 2 years postoperatively, no significant changes were observed in the radiographic images. Lessons: This case report presents an unexpected failure of THA due to superior rim fracture and excessive wear at the locking mechanism of the vitamin E-infused highly cross-linked polyethylene liner. This is an interesting case, as early PE liner failure occurred without strong labor intensity or trauma. A modular component exchange was performed with a dual-mobility bearing, and no issues were observed approximately 2 years after the reoperation. Therefore, third-generation highly cross-linked polyethylene liners can also cause early failure without a clear cause, and this case report highlights the necessity of considering several strategies for reoperation. Study Design: Case report.