z-logo
open-access-imgOpen Access
Numerical Investigation of the Influence of Blade Radial Gap Flow on Axial Blood Pump Performance
Author(s) -
Guang-Mao Liu,
Donghai Jin,
Jianye Zhou,
Yan Zhang,
Haibo Chen,
Hansong Sun,
Shengshou Hu,
Xingmin Gui
Publication year - 2019
Publication title -
asaio journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.961
H-Index - 66
eISSN - 1538-943X
pISSN - 1058-2916
DOI - 10.1097/mat.0000000000000745
Subject(s) - materials science , shroud , mechanics , shear stress , particle image velocimetry , vortex , structural engineering , turbulence , composite material , engineering , physics
The gaps between the blades and the shroud (or hub) of an axial blood pump affect the hydraulics, efficiency, and hemolytic performance. These gaps are critical parameters when a blood pump is manufactured. To evaluate the influence of blade gaps on axial blood pump performance, the flow characteristics inside an axial blood pump with different radial blade gaps were numerically simulated and analyzed with special attention paid to the hydraulic characteristics, gap flow, hydraulic efficiency, and hemolysis index (HI). In vitro hydraulic testing and particle image velocimetry testing were conducted to verify the numerical results. The simulation results showed that the efficiency and pressure rise decreased when the gap increased. The efficiency of the axial blood pump at design point decreased from 37.1% to 27.1% and the pressure rise decreased from 127.4 to 71.2 mm Hg when the gap increased from 0.1 to 0.3 mm. Return and vortex flows were present in the outlet guide vane channels when the gap was larger than 0.2 mm. The HI of the blood pump with a 0.1 mm gap was 1.5-fold greater than that with a 0.3 mm gap. The results illustrated poor hydraulic characteristics when the gap was larger than 0.15 mm and rapidly deteriorated hemolysis when the gap was larger than 0.1 mm. The numerical and experimental results demonstrated that the pressure rise, pump efficiency, and scalar shear stress decreased when the gap increased. The HI did not strictly decrease with gap increases. The preliminary results encourage the improvement of axial blood pump designs.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here