Variations of Intraocular Pressure Measured by Goldmann Applanation Tonometer, Tono-Pen, iCare Rebound Tonometer, and Pascal Dynamic Contour Tonometer in Patients With Corneal Edema After Phacoemulsification
Author(s) -
Weerawat Kiddee,
Arunee Tanjana
Publication year - 2020
Publication title -
journal of glaucoma
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.11
H-Index - 88
eISSN - 1536-481X
pISSN - 1057-0829
DOI - 10.1097/ijg.0000000000001725
Subject(s) - medicine , ophthalmology , goldmann applanation tonometer , intraocular pressure , phacoemulsification , glaucoma , visual acuity
Edematous corneas had lower Goldmann applanation tonometer (GAT) intraocular pressure (IOP) compared with other tonometers. A significant, mild negative correlation between central corneal thickness (CCT) and GAT IOP was found in a group of significant edematous corneas with CCT increases of >100 µm.