Frequency-to-Place Mismatch: Characterizing Variability and the Influence on Speech Perception Outcomes in Cochlear Implant Recipients
Author(s) -
Michael W Canfarotta,
Margaret T. Dillon,
Emily Buss,
Harold C. Pillsbury,
Kevin D. Brown,
Brendan P. O’Connell
Publication year - 2020
Publication title -
ear and hearing
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.577
H-Index - 109
eISSN - 1538-4667
pISSN - 0196-0202
DOI - 10.1097/aud.0000000000000864
Subject(s) - cochlear implant , spiral ganglion , audiology , electrode array , speech perception , hearing aid , acoustics , perception , medicine , physics , hearing loss , electrode , psychology , neuroscience , quantum mechanics
The spatial position of a cochlear implant (CI) electrode array affects the spectral cues provided to the recipient. Differences in cochlear size and array length lead to substantial variability in angular insertion depth (AID) across and within array types. For CI-alone users, the variability in AID results in varying degrees of frequency-to-place mismatch between the default electric frequency filters and cochlear place of stimulation. For electric-acoustic stimulation (EAS) users, default electric frequency filters also vary as a function of residual acoustic hearing in the implanted ear. The present study aimed to (1) investigate variability in AID associated with lateral wall arrays, (2) determine the subsequent frequency-to-place mismatch for CI-alone and EAS users mapped with default frequency filters, and (3) examine the relationship between early speech perception for CI-alone users and two aspects of electrode position: frequency-to-place mismatch and angular separation between neighboring contacts, a metric associated with spectral selectivity at the periphery.