Lesion Formation in Radiofrequency Surgery of the Tongue Base
Author(s) -
Stuck Boris A.,
Köpke Julian,
Maurer Joachim T.,
Verse Thomas,
Eckert Andreas,
Bran Gregor,
Düber Christoph,
Hörmann Karl
Publication year - 2003
Publication title -
the laryngoscope
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.181
H-Index - 148
eISSN - 1531-4995
pISSN - 0023-852X
DOI - 10.1097/00005537-200309000-00030
Subject(s) - tongue , lesion , in vivo , medicine , magnetic resonance imaging , nuclear medicine , biomedical engineering , surgery , pathology , radiology , biology , microbiology and biotechnology
Objectives Temperature controlled radiofrequency volumetric tissue reduction (RFVTR) of the tongue base is a minimally invasive technique for the treatment of obstructive sleep apnea. But despite its widespread use, little is yet known about in vivo effects in humans. Such knowledge would be essential for evidence‐based criteria in the selection of energy application. Methods In a preparatory in vitro study, porcine tongues were preserved in growing medium. Lesions with different amounts of energy were applied, and maximum diameters were assessed. In the in vivo study, 11 patients were treated with RFVTR at the tongue base by employing different energy levels (800, 600, 400, or 200 J) on two application sites. Magnetic resonance imagery scans were performed 4 to 6, 8 to 10, and 24 hours after surgery. An inversion recovery technique was used to visualize the lesions. Results RFVTR created lesions at the porcine tongue from 50 J and higher. Maximum lesion sizes were achieved with 400 J. In vivo, all the lesions were clearly visible in the postoperative scans. Lesion size increased with the amount of energy applied. Maximum diameters were created from 600 J and higher. Higher amounts of energy only resulted in a slight increase in lesion length. Conclusions The application of 600 J at 85°C leads to optimal lesion sizes. Higher amounts of energy will not lead to a relevant increase in tissue necrosis. With regard to the time needed for application, 600 J appears to be the ideal adjustment for energy delivery in the treatment of the human tongue.