z-logo
open-access-imgOpen Access
Silent burn: the hidden danger and effects of bright light from fibre-optic cables in arthroscopic knee surgery
Author(s) -
Jai Chitnavis
Publication year - 2020
Publication title -
journal of surgical case reports
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.14
H-Index - 3
ISSN - 2042-8812
DOI - 10.1093/jscr/rjaa068
Subject(s) - medicine , surgery , perforation , lesion , materials science , punching , metallurgy
Following an uneventful arthroscopic menisectomy of the right knee, a white circular skin lesion, 1 cm in diameter, was noted on the anterior left thigh of a 23-year-old patient. The overlying paper surgical drape had not ignited nor produced smoke. Close inspection revealed a minute perforation in the drape with slight discolouration. No electro-cautery, radio-ablation or irritant skin preparation had been used during surgery. Tests failed to identify fault with the light source, fibre-optic cable or arthroscope. The lesion was diagnosed as a full-thickness thermal burn resulting from heat transmitted from a 300-W Xenon lamp via a detached fibre-optic cable. The effects of contact between an illuminated fibre-optic light cable and living human skin are described, with changes in appearances followed over 2 years. Patients may be burnt and permanently scarred without the knowledge of staff in operating theatres if detached light cables rest against surgical drapes.