Comparison of calibration of surface contamination monitors with Brazilian Network – 2018/2019
Author(s) -
TS Cabral,
Álvaro Laranjeira,
M.P.A. Potiens,
Cma Soares,
Helen J. Khoury,
V Saito,
M David,
Luciano Vidal,
MA Pires
Publication year - 2021
Publication title -
journal of physics. conference series
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.21
H-Index - 85
eISSN - 1742-6596
pISSN - 1742-6588
DOI - 10.1088/1742-6596/1826/1/012039
Subject(s) - calibration , contamination , environmental science , metrology , ionizing radiation , remote sensing , physics , geography , irradiation , statistics , mathematics , optics , nuclear physics , ecology , biology
This work reports the results obtained from the comparison involving 7 laboratories in Brazil. This exercise is about the calibration service of surface contamination monitors. The monitor has been calibrated to BS ISO 7503-3, the calibration factor in terms of surface emission rate. The comparison was conducted by the Brazilian National Laboratory of Ionizing Radiation Metrology (LNMRI / IRD) from October 2018 to July 2019. The extensive sources used were 14 C, 137 Cs, 60 Co, 90 Sr/ 90 Y, 36 Cl and 241 Am. The result of this proficiency test was excellent and proved the calibration capacity of the Brazilian network in the calibration service for surface contamination monitors.