Novel, clinically applicable method to measure step-width during the swing phase of gait
Author(s) -
Tomas Bäcklund,
Fredrik Öhberg,
Gustav Johansson,
Helena Grip,
Nina Sundström
Publication year - 2020
Publication title -
physiological measurement
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.674
H-Index - 101
eISSN - 1361-6579
pISSN - 0967-3334
DOI - 10.1088/1361-6579/ab95ed
Subject(s) - stride , gait , swing , gait analysis , simulation , physical medicine and rehabilitation , computer science , medicine , physics , acoustics
Objective : Step-width during walking is an indicator of stability and balance in patients with neurological disorders, and development of objective tools to measure this clinically would be a great advantage. The aim of this study was to validate an in-house-developed gait analysis system (Striton), based on optical and inertial sensors and a novel method for stride detection, for measuring step-width during the swing phase of gait and temporal parameters. Approach : The step-width and stride-time measurements were validated in an experimental setup, against a 3D motion capture system and on an instrumented walkway. Further, test-retest and day-to-day variability were evaluated, and gait parameters were collected from 87 elderly persons (EP) and four individuals with idiopathic normal pressure hydrocephalus (iNPH) before/after surgery. Main results : Accuracy of the step-width measurement was high: in the experimental setup mean error was 0.08 ± 0.25 cm (R = 1.00) and against the 3D motion capture system 0.04 ± 1.12 cm (R = 0.98). Test-retest and day-to-day measurements were equal within ±0.5 cm. Mean difference in stride time was −0.003 ± 0.008 s between Striton and the instrumented walkway. The Striton system was successfully applied in the clinical setting on individuals with iNPH, which had larger step-width (6.88 cm, n = 4) compared to EP (5.22 cm, n = 87). Significance : We conclude that Striton is a valid, reliable and wearable system for quantitative assessment of step-width and temporal parameters during gait. Initial measurements indicate that the newly defined step-width parameter differs between EP and patients with iNPH and before/after surgery. Thus, there is potential for clinical applicability in patients with reduced gait stability.