Influence of Bolt Tightening Torque, Wire Size, and Component Reuse on Wire Fixation in Circular External Fixation
Author(s) -
Wosar Marc A.,
MarcellinLittle Denis J.,
Roe Simon C.
Publication year - 2002
Publication title -
veterinary surgery
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.652
H-Index - 79
eISSN - 1532-950X
pISSN - 0161-3499
DOI - 10.1053/jvet.2002.34672
Subject(s) - slippage , torque , fixation (population genetics) , tension (geology) , structural engineering , composite material , stainless steel wire , biomedical engineering , materials science , medicine , ultimate tensile strength , engineering , population , physics , environmental health , thermodynamics
Objectives— To evaluate the effects of bolt torque, wire size, and component reuse on the ability to maintain wire tension in 3 external skeletal fixation systems. Study— Design Biomechanical study. Methods— Yield strength in tension of 1.0‐, 1.2‐, 1.5‐, and 1.6‐mm‐diameter wires, and yield strength in torque of Hofmann Small Bone Fixation (SBF) cannulated and slotted bolts and IMEX regular and miniature bolts were determined on a testing machine. The minimum bolt tightening torque needed to prevent wire slippage at clinically recommended wire tensions was determined. Components were tested 10 times, and loads at slippage were recorded. Results— The IMEX system required a mean of 8 Nm of bolt tightening torque to maintain 900 N (1.6‐mm wires). The SBF system required a mean of 3 Nm bolt torque to maintain 300 N (1.0‐mm wires) and 5 Nm to maintain 600 N (1.2‐mm wires). The SBF cannulated bolt required 9 Nm of torque to maintain 900 N (1.5‐mm wires). The SBF slotted bolts could only maintain 800 N before yield. The IMEX miniature system required a mean bolt torque of 1.1 Nm to maintain 300 N. The cannulated and slotted bolts from both manufacturers failed to maintain 70% of initial wire tension after 7 and 4 uses, respectively. Conclusions— The IMEX systems and the SBF system using 1.0‐ and 1.2‐mm wires could maintain clinically recommended wire tension safely. Only the IMEX system could maintain clinically recommended wire tension safely using 1.5‐ or 1.6‐mm wires. Clinical Relevance— The SBF system using 1.0‐ and 1.2‐mm wires and the IMEX system using all wire sizes can maintain clinically relevant wire tension. The SBF system using 1.5‐mm wires could not. Cannulated and slotted bolts should not be used more than 6 and 3 times, respectively. Nuts should not be reused.