Monitoring of radiofrequency thermal ablation in liver tissue through fibre Bragg grating sensors array
Author(s) -
Tosi D.,
Macchi E.G.,
Braschi G.,
Gallati M.,
Cigada A.,
Poeggel S.,
Leen G.,
Lewis E.
Publication year - 2014
Publication title -
electronics letters
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.375
H-Index - 146
eISSN - 1350-911X
pISSN - 0013-5194
DOI - 10.1049/el.2014.0620
Subject(s) - fiber bragg grating , materials science , thermal ablation , thermal , ablation , optics , radiofrequency ablation , grating , liver tissue , optical fiber , optoelectronics , biomedical engineering , physics , engineering , medicine , meteorology , aerospace engineering
A fibre Bragg grating (FBG) array‐based sensing probe has been installed on a device for radiofrequency thermal ablation (RFTA). The probe is made of five FBGs with 0.5 cm active area and 1 cm spacing, to provide quasi‐distributed thermal pattern measurements. Multiple experiments have been conducted on porcine liver, reporting the temperature pattern along the ablation axis. Thermal maps allow an estimation of the RFTA efficiency and spatial extension in the liver.