z-logo
Premium
Implantable Cardioverter‐Defibrillator Malfunction due to Mechanical Failure of the Header Connection
Author(s) -
PICKETT R. ANDREW,
SAAVEDRA PABLO,
ALI MUHAMMED F.,
DARBAR DAWOOD,
ROTTMAN JEFFREY N.
Publication year - 2004
Publication title -
journal of cardiovascular electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.193
H-Index - 138
eISSN - 1540-8167
pISSN - 1045-3873
DOI - 10.1046/j.1540-8167.2004.03682.x
Subject(s) - medicine , header , implantable cardioverter defibrillator , intervention (counseling) , cardiology , computer security , computer science , computer network , psychiatry
Four cases of ICD system malfunction occurring within the Medtronic Marquis family over a limited time frame are reported. These cases shared a common element of defective connections between the header post and the right ventricular pace/sense terminal pin that required intervention and reconnection. Inappropriate shocks occurred in two cases. The circumstances of these cases provide important information about the potential difficulties with these devices and highlight the deficiencies of the current system in tracking and analyzing device‐related problems.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here