Implantable Cardioverter Defibrillator Sensing Failure Due to Endocardial R Wave Electrical Alternans
Author(s) -
JONGNARANGSIN KRIT,
MELA THEOFANIE
Publication year - 2002
Publication title -
journal of cardiovascular electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.193
H-Index - 138
eISSN - 1540-8167
pISSN - 1045-3873
DOI - 10.1046/j.1540-8167.2002.00702.x
Subject(s) - medicine , implantable cardioverter defibrillator , cardiology , defibrillation , ventricular tachycardia , t wave alternans , defibrillation threshold , tachycardia , ventricular fibrillation , sudden cardiac death
Electrical Alternans and ICD Undersensing. A 71‐year‐old patient underwent routine single‐chamber implantable cardioverter defibrillator (ICD) generator replacement. During defibrillation threshold testing, ventricular tachycardia was induced but the ICD failed twice to properly detect the tachycardia due to endocardial R wave alternans and sensing of every other beat. The problem was resolved by inserting a separate sensing/pacing lead positioned further away from the existing ICD lead.