z-logo
Premium
A Centrifugal Pump Driven Tidal Flow Extracorporeal Membrane Oxygenation System Tested with Neonatal Mock Circulation
Author(s) -
Trittenwein Gerhard,
Kölbl Reinhard,
Trittenwein Hildegard,
Golej Johann,
Burda Gudrun,
Hermon Michael,
Pollak Arnold
Publication year - 1999
Publication title -
artificial organs
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.684
H-Index - 76
eISSN - 1525-1594
pISSN - 0160-564X
DOI - 10.1046/j.1525-1594.1999.06395.x
Subject(s) - extracorporeal membrane oxygenation , centrifugal pump , medicine , hemodynamics , extracorporeal circulation , cannula , circulatory system , peristaltic pump , anesthesia , cardiology , mechanics , surgery , engineering , impeller , physics , mechanical engineering
In 1993, Chevalier published his experiences with tidal flow venovenous extracorporeal membrane oxygenation (ECMO) featuring a single lumen cannula, nonocclusive roller pump, and alternating clamps. Using a neonatal mock circulation (NMC), which enables different hemodynamic states for neonatal ECMO research, the tested hypothesis was that it is possible to create a centrifugal pump driven tidal flow neonatal venovenous ECMO system. Additionally, the resulting hemodynamic effects in a condition of circulatory impairment were investigated. The ECMO circuit tested was assembled using a pediatric centrifugal pump head, a distensible reservoir, and a rotary clamp separating drainage from the injection phase. Using the NMC, end tidal volumes, mock circulation flow, and arterial and venous pressures were measured at different pump speeds after the drainage and injection phases. Effective venovenous ECMO flow (evvEF) was calculated. Mock circulation baseline values (ECMO clamped) were compared to values during tidal flow ECMO. At 3,000 rpm, a centrifugal pump speed of 75 ml/kg/min evvEF was reached, and it increased with higher pump speeds. At this point, the end tidal mock circulation flow (representing cardiac output) after drainage differed significantly from that during the injection phase (p < 0.01) but not from the baseline value. The end tidal arterial and venous pressures after the drainage phase were found to be significantly decreased compared to the baselines (p < 0.01). In conclusion, a centrifugal pump driven tidal flow venovenous ECMO system can be created enabling sufficient tidal volumes. Tested in the described NMC simulating posthypoxic circulatory impairment, significant hemodynamic effects could be demonstrated. Animal experiments for confirmation are necessary.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here