Inhibition of Biventricular Pacemakers by Oversensing of Far‐Field Atrial Depolarization
Author(s) -
LIPCHENCA IGOR,
GARRIGUE STÉPHANE,
GLIKSON MICHAEL,
BAROLD S. SERGE,
CLÉMENTY JACQUES
Publication year - 2002
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1046/j.1460-9592.2002.00365.x
Subject(s) - medicine , cardiology , heart failure , ventricular pacing
LIPCHENCA, I., et al. : Inhibition of Biventricular Pacemakers by Oversensing of Far‐Field Atrial Depo‐larization. This report describes two patients who exhibited far‐field oversensing of the P wave by the ventricular channel of a DDD biventricular pacemaker implanted for the treatment of congestive heart failure. Oversensing probably occurred secondary to slight displacement of the left ventricular lead in the coronary venous system. Long‐term reliable pacing was restored by decreasing the sensitivity of the ventricular channel.