Premium
Pacemaker‐Mediated Tachycardia in a Biventricular Pacing System
Author(s) -
VAN GELDER BERRY M.,
BRACKE FRANK A.,
MEIJER ALBERT
Publication year - 2001
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1046/j.1460-9592.2001.01819.x
Subject(s) - medicine , cardiology , ventricular pacing , ventricular tachycardia , tachycardia , atrial fibrillation , heart failure , anesthesia
VAN GELDER, B.M., et al. ; Pacemaker‐Mediated Tachycardia in a Biventricular Pacing System. A 63‐year‐old man with chronic atrial fibrillation and heart failure had a biventricular pacing system implanted. The pulse generator was a standard DDDR pacemaker, using the atrial channel for the right ventricular lead and the ventricular channel for the left ventricular lead. During final adjustment of the pacing parameters, a pacemaker tachycardia triggered by T wave oversensing from the right ventricular lead was recorded.