Paradoxical Induction of Cross‐Talk in a Dual Chamber Pacemaker by Increasing the Ventricular Output
Author(s) -
BAROLD S. SERGE
Publication year - 2001
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1046/j.1460-9592.2001.00906.x
Subject(s) - medicine , cardiology , stimulus (psychology) , ventricular pacing , heart failure , cognitive psychology , psychology
BAROLD, S.S.: Paradoxical Induction of Cross‐Talk in a Dual Chamber Pacemaker by Increasing the Ventricular Output. Cross‐talk was induced in a unipolar DDD pacemaker by increasing the output of the ventricular channel. Oversensing by the ventricular channel was probably due to the detection of a relatively large summation signal consisting of the afterpotential and the superimposed activity from the atrial stimulus.