The effect of large solid angles of collection on quantitative X‐ray microanalysis in the AEM
Author(s) -
Toru Watanabe,
Ackland,
Michelle D. Williams
Publication year - 1999
Publication title -
journal of microscopy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.569
H-Index - 111
eISSN - 1365-2818
pISSN - 0022-2720
DOI - 10.1046/j.1365-2818.1999.00481.x
Subject(s) - microanalysis , solid angle , analytical chemistry (journal) , absorption (acoustics) , optics , range (aeronautics) , scanning electron microscope , x ray , scanning transmission electron microscopy , materials science , microscope , detector , chemistry , physics , composite material , organic chemistry , chromatography
Increasing the solid angle of X‐ray collection is a major factor in improving the analytical sensitivity of X‐ray energy‐dispersive spectrometry (XEDS) in the analytical electron microscope (AEM). A new scanning transmission electron microscope, the VG HB 603, is equipped with two XEDS detectors with the largest collection angles (0.30 and 0.17 sr) available in commercial AEMs. However, large collection angles result in a large range of take‐off angles, from ~ 4° to 36°, and the low angles can cause strong X‐ray absorption. In order to investigate possible detrimental effects of the low (and of the range of) take‐off angles on quantitative microanalysis of specimens exhibiting significant absorption, a stoichiometric Ni 3 Al thin‐film, in which the Al Kα line is significantly absorbed, was analysed. Furthermore, the effect of different values of the collection angle on X‐ray intensities was theoretically evaluated by numerical calculations and spectral simulation. These theoretical approaches permitted correlation of changes in the X‐ray take‐off angle (and hence X‐ray absorption) with changes in the collection angle. It is demonstrated that ~ 0.30 sr detectors, with minimum take‐off angles as small as 4°, only result in maximum errors of 4% in the quantification of Al in Ni 3 Al and, therefore, further increases in collection angle can be pursued while maintaining current levels of accuracy of quantification.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here