Differential Ion Mobility Spectrometry Coupled to Tandem Mass Spectrometry Enables Targeted Leukemia Antigen Detection
Author(s) -
Udara Dharmasiri,
Samantha L. Isenberg,
Gary L. Glish,
Paul M. Armistead
Publication year - 2014
Publication title -
journal of proteome research
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.644
H-Index - 161
eISSN - 1535-3907
pISSN - 1535-3893
DOI - 10.1021/pr500527c
Subject(s) - mass spectrometry , chemistry , tandem mass spectrometry , chromatography , analyte , antigen , peptide , tandem , analytical chemistry (journal) , materials science , biochemistry , biology , genetics , composite material
Differential ion mobility spectrometry (DIMS) can be used as a filter to remove undesired background ions from reaching the mass spectrometer. The ability to use DIMS as a filter for known analytes makes DIMS coupled to tandem mass spectrometry (DIMS-MS/MS) a promising technique for the detection of cancer antigens that can be predicted by computational algorithms. In experiments using DIMS-MS/MS that were performed without the use of high-performance liquid chromatography (HPLC), a predicted model antigen, GLR (FLSSANEHL), was detected at a concentration of 10 pM (20 amol) in a mixture containing 94 competing model peptide antigens, each at a concentration of 1 μM. Without DIMS filtering, the GLR peptide was undetectable in the mixture even at 100 nM. Again, without using HPLC, DIMS-MS/MS was used to detect 2 of 3 previously characterized antigens produced by the leukemia cell line U937.A2. Because of its sensitivity, a targeted DIMS-MS/MS methodology can likely be used to probe for predicted cancer antigens from cancer cell lines as well as human tumor samples.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here