Greater than 3-Log Reduction in Viable Coronavirus Aerosol Concentration in Ducted Ultraviolet-C (UV–C) Systems
Author(s) -
Yuechen Qiao,
My Yang,
Ian A. Marabella,
Devin A.J. McGee,
Hamada A. Aboubakr,
Sagar M. Goyal,
Christopher J. Hogan,
Bernard A. Olson,
Montserrat Torremorell
Publication year - 2020
Publication title -
environmental science and technology
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 2.851
H-Index - 397
eISSN - 1520-5851
pISSN - 0013-936X
DOI - 10.1021/acs.est.0c05763
Subject(s) - aerosol , ultraviolet , covid-19 , coronavirus , reduction (mathematics) , environmental science , ultraviolet radiation , chemistry , environmental chemistry , virology , materials science , medicine , optoelectronics , radiochemistry , mathematics , organic chemistry , infectious disease (medical specialty) , geometry , outbreak , disease , pathology
Control technologies to inactivate airborne viruses effectively are needed during the ongoing SARS-CoV-2 pandemic, and to guard against airborne transmitted diseases. We demonstrate that sealed UV-C flow reactors operating with fluences near 253 ± 1 nm of 13.9-49.6 mJ cm -2 efficiently inactivate coronaviruses in an aerosol. For measurements, porcine respiratory coronavirus (PRCV) was nebulized in a custom-built, 3.86 m wind tunnel housed in a biosafety level class II facility. The single pass log 10 reduction of active coronavirus was in excess of 2.2 at a flow rate of 2439 L min -1 (13.9 mJ cm -2 ) and in excess of 3.7 (99.98% removal efficiency) at 684 L min -1 (49.6 mJ cm -2 ). Because virus titers resulting from sampling downstream of the UV-C reactor were below the limit of detection, the true log reduction is likely even higher than measured. Comparison of virus titration results to reverse transcriptase quantitative PCR and measurement of fluorescein concentrations (doped into the nebulized aerosol) reveals that the reduction in viable PRCV is primarily due to UV-C based inactivation, as opposed to physical collection of virus. The results confirm that UV-C flow reactors can efficiently inactivate coronaviruses through incorporation into HVAC ducts or recirculating air purifiers.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom