The role of AV and VV optimization for CRT
Author(s) -
Brabham William W.,
Gold Michael R.
Publication year - 2013
Publication title -
journal of arrhythmia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.463
H-Index - 21
eISSN - 1883-2148
pISSN - 1880-4276
DOI - 10.1016/j.joa.2013.02.001
Subject(s) - medicine , cardiac resynchronization therapy , cardiology , ventricular function , heart failure , ejection fraction
Cardiac resynchronization therapy is an effective therapy for patients with left ventricular systolic dysfunction and a ventricular conduction delay; however, approximately 30% of patients do not experience significant clinical improvement with this treatment. Modern devices allow individualized programming of the AV delay and VV offset, which offer the possibility of improving clinical response rates with optimized programming. AV and VV delay optimization techniques have included echocardiography, device‐based algorithms, and several other novel noninvasive techniques. While an acute improvement in hemodynamic function has been clearly demonstrated with optimized device settings, long‐term clinical benefit is limited. In the majority of cases, an empiric AV delay with simultaneous biventricular or left ventricular pacing is adequate. The value of optimization of these intervals in “non‐responders” still requires further investigation.