Twiddler's syndrome detected by patient's complaint of implantable cardioverter‐defibrillator rotation in the subcutaneous pocket
Author(s) -
Konishi Hirokazu,
Tokano Takashi,
Nakazato Yuji,
Komatsu Sayaka,
Suwa Satoru,
Komatsu Kaoru,
Hayashi Hidemori,
Sekita Gaku,
Sumiyoshi Masataka,
Bito Fumiyasu,
Kizu Kyoko,
Daida Hiroyuki
Publication year - 2012
Publication title -
journal of arrhythmia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.463
H-Index - 21
eISSN - 1883-2148
pISSN - 1880-4276
DOI - 10.1016/j.joa.2011.12.003
Subject(s) - medicine , implantable cardioverter defibrillator , fluoroscopy , surgery , cardiology
Twiddler's syndrome presents as a rare implantable‐cardioverter defibrillator (ICD) malfunction that is potentially fatal. The following case is that of a 79‐year‐old woman, who was implanted with an ICD and experienced Twiddler's syndrome. Five months after the implantation, the patient complained that the generator sometimes rotated in the subcutaneous pocket. Fluoroscopy showed that the generator had rotated and the leads were twisted, which is compatible with Twiddler's syndrome. Although all the leads were functioning normally, a procedure to repair the twisted leads was performed. Twiddler's syndrome is usually detected due to a serious device malfunction; however, this case study suggests that it can be detected prior to ICD malfunction by the patient's complaint.