Averaging of results derived from different, simultaneously acquired mass transitions in ID-LC-MS/MS – Potential impact on measurement imprecision
Author(s) -
Carina Schuster,
Katharina Habler,
Michael Vogeser
Publication year - 2020
Publication title -
clinical mass spectrometry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.473
H-Index - 8
eISSN - 2376-9998
pISSN - 2213-8005
DOI - 10.1016/j.clinms.2020.06.001
Subject(s) - analyte , mass spectrometry , chemistry , chromatography , analytical chemistry (journal) , biological system , mathematics , biology
LC-MS/MS allows for many measurands monitoring different mass transitions simultaneously. So far, such alternative mass transitions are usually assessed as "quantifier and qualifier ions" in order to rule out interferences in individual samples. However, quantification can also be based on assessment of alternative mass transitions for both the measurand and its internal standard, with two distinct results for one injection of an individual sample. These paired results can be averaged. The aim of this study was to determine the potential impact of this averaging approach on measurement imprecision.