Premium
Guideline Summary: Processing Flexible Endoscopes
Author(s) -
AORN
Publication year - 2016
Publication title -
aorn journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.222
H-Index - 43
eISSN - 1878-0369
pISSN - 0001-2092
DOI - 10.1016/j.aorn.2016.06.004
Subject(s) - guideline , citation , computer science , library science , medicine , pathology
III Flexible endoscopes and accessories should be precleaned at the point of use. Precleaning of flexible endoscopes and accessories at the point of use should occur as soon as possible after the endoscope is removed from the patient (or the procedure is completed) and before organic material has dried on the surface or in the channels of the endoscope. When the precleaning process will be delayed (eg, an endoscope is used for intubation and remains in the procedure room for potential reuse), designated personnel (eg, RN circulator, scrub person) should wipe the external surfaces with a soft, lint-free cloth or sponge saturated with utility or sterile water and should suction water through the channels.