z-logo
open-access-imgOpen Access
Cobalt–Carbon Eutectic Fixed Point for Contact Thermometry
Author(s) -
A. D. W. Todd,
M. Gotoh,
D. Woods,
Kathryn Hill
Publication year - 2010
Publication title -
international journal of thermophysics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.32
H-Index - 68
eISSN - 1572-9567
pISSN - 0195-928X
DOI - 10.1007/s10765-010-0907-5
Subject(s) - eutectic system , thermocouple , materials science , calibration , cobalt , fixed point , analytical chemistry (journal) , composite material , metallurgy , microstructure , chemistry , chromatography , mathematics , mathematical analysis , statistics
Two Co\u2013C eutectic fixed points were constructed for thermocouple calibration. The eutectic fixed points were measured with a Pt/Pd thermocouple calibrated at the freezing temperatures of Sn, Zn, Al, Ag, and Au. A temperature of (1323.99 \ub1 0.52)\u25e6C(k = 2) was determined via thismethod. The cell design allowed filling to be accomplished in a single step. Each cellwas held above 1300 \u25e6C for at least 42 h and was subjected to at least 20 melt/freeze cycles with no mechanical failure occurring.abstractPeer reviewed: YesNRC publication: Ye

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom