Photon‐excited K X‐ray fluorescence cross‐section measurements for some low‐ Z ‐elements
Author(s) -
Rani Anita,
Koshal Rajesh K.,
Chaturvedi S. N.,
Nath N.
Publication year - 1988
Publication title -
x‐ray spectrometry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.447
H-Index - 45
eISSN - 1097-4539
pISSN - 0049-8246
DOI - 10.1002/xrs.1300170205
Subject(s) - fluorescence , photoelectric effect , excited state , x ray fluorescence , atomic physics , excitation , cross section (physics) , torr , analytical chemistry (journal) , photon , x ray , resolution (logic) , chemistry , physics , optics , quantum mechanics , chromatography , artificial intelligence , computer science , thermodynamics
X‐ray fluorescence cross‐section for Kα x‐ray lines were measured for the light elements Na, Mg, Al, Si, S, Cl and K relative to Ca at a photon excitation energy of 5.97 keV (Mn K x‐ray). A windowless retractable Kevex Si(Li) detector system having an energy resolution of 146 eV at 6.4 keV was used for the detection of the characteristic K x‐rays. The measurements were performed in a vacuum chamber maintained at a pressure of ca 10 −7 Torr using a 55 Fe radioactive excitation source and targets of ultra‐pure elements/compounds of the above‐mentioned elements. Cross‐sections were also calculated using photoelectric cross‐sections, fluorescence yields and the fractional x‐ray emission rates. Measured and calculated x‐ray fluorescence cross‐sections agreed within reasonable limits.