Summary Measurement of fast pressure changes in the rough vacuum range Using commercially available vacuum gauges, pressure jumps executed in the millisecond range between 1000 and just below 1 mbar were measured in a specially developed system. Of the investigated vacuum gauges, the piezoresistive devices had the shortest response times. They reacted within 1 – 5 ms and achieved a valid measured value after approximately 20 ms. The capacitive manometers responded after 5 – 20 ms and reached a valid measured value after 50 – 100 ms. Lower‐priced Pirani gauges reacted with a response time of 50 – 100 ms and a valid measured value after 200 – 500 ms, which is based on the measuring principle. Besides, a pressure measurement opposed to the actual pressure tendency can be problematic with Piranis.