Determination of 8‐epi PGF 2 α concentrations as a biomarker of oxidative stress using triple‐stage liquid chromatography/tandem mass spectrometry
Author(s) -
Bastani Nasser E.,
Gundersen Thomas E.,
Blomhoff Rune
Publication year - 2009
Publication title -
rapid communications in mass spectrometry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.528
H-Index - 136
eISSN - 1097-0231
pISSN - 0951-4198
DOI - 10.1002/rcm.4197
Subject(s) - chemistry , chromatography , selected reaction monitoring , tandem mass spectrometry , isoprostanes , electrospray ionization , mass spectrometry , detection limit , high performance liquid chromatography , liquid chromatography–mass spectrometry , analyte , isoprostane , urine , oxidative stress , lipid peroxidation , biochemistry
F2‐isoprostanes are a family of prostaglandin F2‐like compounds that are formed by free‐radical‐catalyzed peroxidation of arachidonic acid. Several F2‐isoprostanes, but in particular 8‐epi PGF 2 α , are widely used as oxidative stress biomarkers. An analytical method based on liquid chromatography with negative electrospray ionization (ESI) coupled to tandem mass spectrometric detection (LC/MS/MS) was developed for the determination of 8‐epi PGF 2 α concentrations in human plasma, whole blood, erythrocytes and urine. 8‐epi PGF 2 α ‐d4, a stable isotope derivative of 8‐epi PGF 2 α , was used as an internal standard (IS). A 50 µL sample was focused on‐column and separated on two 3 µm particle size SUPELCOSIL™ ABZ+Plus HPLC columns (15 cm × 4.6 mm and 7.5 cm × 4.6 mm) connected in series. An Applied Biosystems 4000 Q TRAP LC/MS/MS system with ESI was operated in multiple reaction monitoring (MRM) mode with the precursor‐to‐product ion transitions m / z 353.4 → 193.1 (8‐epi PGF 2 α ), 357.4 → 197.1 (8‐epi PGF 2α ‐d4), used for quantification. The assay was fully validated and found to have adequate accuracy, precision, linearity, sensitivity and selectivity. The mass limit of detection (mLOD) was 1 pg of analyte eluting from the column. The assay has been successfully applied to the analysis of human plasma, whole blood, erythrocytes and urine samples. Copyright © 2009 John Wiley & Sons, Ltd.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom