Calibration and mass measurement in negative‐ion fast‐atom bombardment mass spectrometry
Author(s) -
Sim P. G.,
Boyd R. K.,
Bertrand M. J.
Publication year - 1991
Publication title -
rapid communications in mass spectrometry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.528
H-Index - 136
eISSN - 1097-0231
pISSN - 0951-4198
DOI - 10.1002/rcm.1290051111
Subject(s) - chemistry , mass spectrometry , calibration , fast atom bombardment , ion , range (aeronautics) , analytical chemistry (journal) , calibration curve , chromatography , physics , detection limit , materials science , organic chemistry , quantum mechanics , composite material
The problem of acceptable accuracy in mass calibration for fast‐atom bombardment (FAB) mass spectrometry has recently received some attention (K. Véekey, Rapid Commun. Mass Spectrom. , 2, 213 (1988); Idem., Org. Mass Spectrom. , 24, 183 (1989). We describe here a calibration method for negative‐ionFAB, using gold cluster ions, which provides facile mass calibration over the range from can 200 u to 10000 u . Under suitable conditions, it is possible to use these ions as internal standards for accurate mass measurement.