Subject(s) - echo (communications protocol) , nuclear magnetic resonance , fast spin echo , spin echo , physics , magnetic resonance imaging , computer science , medicine , radiology , computer network
A rapid method of simultaneous T 1 and T 2 measurement is presented which uses a segmented echo‐planar readout with varying repetition times (TR) and echo times (TE). This method is useful in T 1 mapping for analysis of dynamic contrast enhanced MRI (DCE‐MRI), where T 1 can be used to estimate contrast agent concentration. In the application of this method to dynamic imaging, the equilibrium magnetization is measured on pre‐contrast images and incorporated into post‐contrast T 1 calculations for improved accuracy. Simultaneous T 2 measurement allows correction of T 2 effects in the T 1 map which may occur at high contrast agent concentrations, and is performed without significant imaging time penalty. Phantom and in vivo results show the usefulness of this technique for analysis of contrast enhancement kinetics. Accurate rapid contrast agent concentration measurement may be useful for analyzing the distribution and kinetics of contrast agents or labeled pharmaceuticals. Copyright © 2009 John Wiley & Sons, Ltd.