A multi spin echo pulse sequence with optimized excitation pulses and a 3D cone readout for hyperpolarized 13 C imaging
Author(s) -
Somai Vencel,
Wright Alan J.,
Fala Maria,
Hesse Friederike,
Brindle Kevin M.
Publication year - 2020
Publication title -
magnetic resonance in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.696
H-Index - 225
eISSN - 1522-2594
pISSN - 0740-3194
DOI - 10.1002/mrm.28248
Subject(s) - pulse sequence , nuclear magnetic resonance , excitation , physics , pulse (music) , spin echo , optics , in vivo , chemistry , magnetic resonance imaging , detector , medicine , microbiology and biotechnology , quantum mechanics , biology , radiology
Purpose Imaging tumor metabolism in vivo using hyperpolarized [1‐ 13 C]pyruvate is a promising technique for detecting disease, monitoring disease progression, and assessing treatment response. However, the transient nature of the hyperpolarization and its depletion following excitation limits the available time for imaging. We describe here a single‐shot multi spin echo sequence, which improves on previously reported sequences, with a shorter readout time, isotropic point spread function (PSF), and better signal‐to‐noise ratio. Methods The sequence uses numerically optimized spectrally selective excitation pulses set to the resonant frequencies of pyruvate and lactate and a hyperbolic secant adiabatic refocusing pulse, all applied in the absence of slice selection gradients. The excitation pulses were designed to be resistant to the effects of B 0 and B 1 field inhomogeneity. The gradient readout uses a 3D cone trajectory composed of 13 cones, all fully refocused and distributed among 7 spin echoes. The maximal gradient amplitude and slew rate were set to 4 G/cm and 20 G/cm/ms, respectively, to demonstrate the feasibility of clinical translation. Results The pulse sequence gave an isotropic PSF of 2.8 mm. The excitation profiles of the optimized pulses closely matched simulations and a 46.10 ± 0.04% gain in image SNR was observed compared to a conventional Shinnar–Le Roux excitation pulse. The sequence was demonstrated with dynamic imaging of hyperpolarized [1‐ 13 C]pyruvate and [1‐ 13 C]lactate in vivo. Conclusion The pulse sequence was capable of dynamic imaging of hyperpolarized 13 C labeled metabolites in vivo with relatively high spatial and temporal resolution and immunity to system imperfections.