Closed‐form solution for T 2 mapping with nonideal refocusing of slice selective CPMG sequences
Author(s) -
Petrovic Andreas,
Scheurer Eva,
Stollberger Rudolf
Publication year - 2015
Publication title -
magnetic resonance in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.696
H-Index - 225
eISSN - 1522-2594
pISSN - 0740-3194
DOI - 10.1002/mrm.25170
Subject(s) - nuclear magnetic resonance , physics , chemistry , materials science
Purpose T 2 quantification with multiecho sequences is typically impaired by the contribution of stimulated echoes to the echo decay due to B 1 + inhomogeneity and slice profile effects. In this work, a compact signal model based on the generating functions approach, which accounts for both sources of error, is presented. Methods The generating functions (GF) approach is used to obtain a closed solution to the evolution of the transverse magnetization in an echo train, however, not in the time domain, but in the transformed z‐domain. The approach is generalized by the incorporation of flip angle distribution across the refocusing slice profiles. The approach is tested by fitting the model to simulated data as well as to phantom and in vivo measurements, followed by a comparison with the common monoexponential fitting approach. Results The fitting simulations indicate that T 2 errors of up to 30% can be commonplace in a clinical setting using the monoexponential method. Conversely, the GF approach produced accurate results. Phantom and in vivo experiments showed a good agreement of the GF values with spectroscopic measurements and single‐echo spin‐echo sequences. Conclusion A correction for stimulated echoes is necessary to compute comparable T 2 values. The presented approach provides a solution to this issue. Magn Reson Med 73:818–827, 2015. © 2014 Wiley Periodicals, Inc.