Fat quantification using multiecho sequences with bipolar gradients: Investigation of accuracy and noise performance
Author(s) -
Peterson Pernilla,
Månsson Sven
Publication year - 2014
Publication title -
magnetic resonance in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.696
H-Index - 225
eISSN - 1522-2594
pISSN - 0740-3194
DOI - 10.1002/mrm.24657
Subject(s) - noise (video) , imaging phantom , sensitivity (control systems) , signal (programming language) , amplitude , echo (communications protocol) , computer science , accuracy and precision , acoustics , electromagnetic coil , artifact (error) , nuclear magnetic resonance , physics , artificial intelligence , optics , electronic engineering , computer network , quantum mechanics , image (mathematics) , programming language , engineering
Purpose To investigate the accuracy and noise performance of fat quantification with multiple gradient‐echo images acquired using bipolar read‐out gradients and compare them with those of the well‐established unipolar technique. Theory The bipolar read‐out technique induces phase and amplitude errors caused by gradient delays, eddy currents, and frequency‐dependent coil sensitivity. In this study, these errors were corrected for jointly with the fat/water separation by modeling the impact of these effects on the signal. This approach did not require acquisition of reference data or modification of the pulse sequence. Methods Simulations and a phantom experiment were used to investigate the accuracy and noise performance of the technique and compare them with those of a well‐established technique using unipolar read‐out gradients. Also, the in vivo feasibility was demonstrated for abdominal applications. Results The phantom experiment demonstrated similar accuracy of the bipolar and unipolar fat quantification techniques. In addition, the noise performance was shown not to be affected by the added estimations of the phase and amplitude errors for most inter‐echo times. Conclusion The bipolar technique was found to provide accurate fat quantification with noise performance similar to the unipolar technique given an appropriate choice of inter‐echo time. Magn Reson Med 71:219–229, 2014. © 2013 Wiley Periodicals, Inc.