Premium
Acquistion and Accuracy in Rapid NMR Imaging Methods
Author(s) -
Twieg Donald B.
Publication year - 1985
Publication title -
magnetic resonance in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.696
H-Index - 225
eISSN - 1522-2594
pISSN - 0740-3194
DOI - 10.1002/mrm.1910020504
Subject(s) - nuclear magnetic resonance , computer science , nuclear medicine , physics , medicine
The echo‐planar imaging method (EPI) and its variants can acquire data sufficient to form an image in a much shorter time than the Fourier or projection‐reconstruction methods. Fundamental requirements on gradient waveforms and signal sampling are discussed for EPI and SEPI (a variant using sinusoidal gradient waveforms), based on the k ‐trajectory formulation of acquisition in NMR imaging. Using a new simulation method, image defects in EPI and SEPI are demonstrated which result from failure to meet these requirements. Faithful rendition of theoretically ideal gradient waveforms places severe demands on gradient system response in rapid imaging. It is shown, however, that nonideal gradient system response need not degrade images if appropriate reconstruction computations are used. Transfer functions and noise power spectra are derived for EPI and SEPI, and the ratio of these two functions is used to derive ideal observer signal‐to‐noise ratios for simple visual interpretive tasks. It is shown that ideal observer performance can differ significantly between EPI and SEPI for some visual tasks, using the same NMR imaging system.© 1985 Academic Press,Inc.