Rapid multicontrast brain imaging on a 0.35T MR‐linac
Author(s) -
NejadDavarani Siamak P.,
Zakariaei Niloufar,
Chen Yongsheng,
Haacke E. Mark,
Hurst Newton J.,
Salim Siddiqui M.,
Schultz Lonni R.,
Snyder James M.,
Walbert Tobias,
GlideHurst Carri K.
Publication year - 2020
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1002/mp.14251
Subject(s) - imaging phantom , linear particle accelerator , nuclear medicine , flip angle , repeatability , magnetic resonance imaging , nuclear magnetic resonance , physics , medicine , mathematics , radiology , beam (structure) , optics , statistics
Purpose Magnetic resonance‐guided radiation therapy (MRgRT) has shown great promise for localization and real‐time tumor monitoring. However, to date, quantitative imaging has been limited for low field MRgRT. This work benchmarks quantitative T1, R2*, and Proton Density (PD)mapping in a phantom on a 0.35 T MR‐linac and implements a novel acquisition method, ST rategically A cquired G radient E cho (STAGE). To further validate STAGE in a clinical setting, a pilot study was undertaken in a cohort of brain tumor patients to elucidate opportunities for longitudinal functional imaging with an MR‐linac in the brain. Methods STAGE (two triple‐echo gradient echo (GRE) acquisitions) was optimized for a 0.35T low‐field MR‐linac. Simulations were performed to choose two flip angles to optimize signal‐to‐noise ratio (SNR) and T1‐mapping precision. Tradeoffs between SNR, scan time, and spatial resolution for whole‐brain coverage were evaluated in healthy volunteers. Data were inputted into a STAGE processing pipeline to yield four qualitative images (T1‐weighted, enhanced T1‐weighted, proton‐density (PD) weighted, and simulated FLuid‐Attenuated Inversion Recovery (sFLAIR)), and three quantitative datasets (T1, PD, and R2*). A benchmarking ISMRM/NIST phantom consisting of vials with variable NiCl 2 and MnCl 2 concentrations was scanned using variable flip angles (VFA) (2–60 degrees) and inversion recovery (IR) methods at 0.35 T. STAGE and VFA T1 values of vials were compared to IR T1 values. As measures of agreement with reference values and repeatability, relative error (RE) and coefficient of variability (CV) were calculated, respectively, for quantitative MR values within the phantom vials (spheres). To demonstrate feasibility, longitudinal STAGE data (pretreatment, weekly, and ~ 2 months post‐treatment) were acquired in an IRB‐approved pilot study of brain tumor cases via the generation of temporal and differential quantitative MRI maps. Results In the phantom, RE of measured VFA T1 and STAGE relative to IR reference values were 7.0 ± 2.5% and 9.5 ± 2.2% respectively. RE for the PD vials was 8.1 ± 6.8% and CV for phantom R2* measurements was 10.1 ± 9.9%. Simulations and volunteer experiments yielded final STAGE parameters of FA = 50°/10°, 1 × 1 × 3 mm 3 resolution, TR = 40 ms, TE = 5/20/34 ms in 10 min (64 slices). In the pilot study of brain tumor patients, differential maps for R2* and T1 maps were sensitive to local tumor changes and appeared similar to 3 T follow‐up MRI datasets. Conclusion Quantitative T1, R2*, and PD mapping are promising at 0.35 T agreeing well with reference data. STAGE phantom data offer quantitative representations comparable to traditional methods in a fraction of the acquisition time. Initial feasibility of implementing STAGE at 0.35 T in a patient brain tumor cohort suggests that detectable changes can be observed over time. With confirmation in a larger cohort, results may be implemented to identify areas of recurrence and facilitate adaptive radiation therapy.