Characterization methods for comprehensive evaluations of shielding materials used in an MRI
Author(s) -
GrossWeege Nicolas,
Dey Thomas,
Gebhardt Pierre,
Schug David,
Weissler Bjoern,
Schulz Volkmar
Publication year - 2018
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1002/mp.12762
Subject(s) - electromagnetic shielding , medical physics , characterization (materials science) , medical imaging , materials science , biomedical engineering , computer science , medicine , radiology , nanotechnology , composite material
Purpose In order to integrate electronic devices into a magnetic resonance imaging (MRI) scanner, shielding of the electronics with respect to the radio frequency (RF) transmit and receive system of the MRI scanner is required. Furthermore, MRI uses time‐varying low‐frequency magnetic fields for spatial encoding, i.e., the gradient magnetic fields. Time‐varying magnetic fields induce eddy currents in all conductive elements. The eddy currents result in opposing magnetic fields, which can cause distortions of the magnetic resonance (MR) image. As shielding of lower frequencies is not feasible in this respect, an ideal shielding element should be transparent for gradient magnetic fields while providing a high RF shielding effectiveness. Furthermore, it should offer a low susceptibility to prevent distortion of the main magnetic field of the MRI. In this work, we characterize the aforesaid shielding parameters of different shielding samples. Methods We developed a nuclear magnetic resonance (NMR) probe to measure the magnetic fields to quantify the field distortions time‐resolvedly. The relative distortion was introduced as a proportionality constant relating the eddy‐current‐inducing field changes and the field distortions. The relative distortion was measured in the frequency range from 0 to 10 kHz for all shielding samples using the NMR probe. We characterized the shielding effectiveness of the samples in the frequency range from 1 to 150 MHz using a network analyzer. We conducted all measurements with three different materials, two carbon fiber composites and copper, each in various thicknesses. Results The relative distortion of the magnetic fields induced by the carbon fiber composites samples was at least a factor of seven lower than the copper sample. A linear dependency on the sample thickness was measured for the main field distortion, the relative distortion and the shielding effectiveness. The relative distortion was roughly independent of the gradient frequency contrary to the shielding effectiveness, highly depending on the RF frequency. Conclusions We presented a very sensitive method to characterize the distortion of the main field distortion and the gradient transparency using an NMR probe. We analyzed different shielding materials regarding the main field distortion, the gradient transparency, and the shielding effectiveness. From the tested materials, we identified a carbon fiber composite with the lowest distortion on the MRI.